Median Gain and Frequency Response Flatness

Specifications (at 25 °C ±5 °C)
Median gain over 300 kHz to 1 GHz
Frequency response flatness (relative to med ian gain)
300 kHz to 1 GHz
1 GHz to 3 GHz
Description
This test procedure performs a response calibration on the analyzer and then
measures the gain of the probe over frequency. Maximum and minimum gains in
two different frequency bands are used to determine the probe performance.
Procedure
1 Preset the analyzer using the [PRESET] hardkey.
2 Set the frequency sweep range for 300 kHz to 3 GHz.
3 Set the analyzer for a transmission (S21) measurement.
4 Connect a type-N cable between the output port and input port on the
5 Perform a response calibration on the analyzer. The trace on the analyzer
6 Disconnect the cable from the output port of the analyzer. Connect the type-N
7 Connect the probe adapter to the output port of the analyzer. Connect the
Keysight 85024A User's and Service Guide
analyzer.
should now be a flat line at 0 dB.
barrel to the free end of the cable.
probe between the probe adapter and the type-N barrel. Plug the probe power
cable into the front panel of the analyzer.
Performance Tests
0 dB ±1.25 dB
±1.25 dB
±2.5 dB