A2 Analog Board Replacement; Tools Required; Removal Procedure; Replacement Procedure - Keysight E4980A Service Manual

Precision lcr meter
Hide thumbs Also See for E4980A:
Table of Contents

Advertisement

Replacement Procedure

A2 Analog Board Replacement

A2 Analog Board Replacement

Tools Required

Removal Procedure

Refer to
Step 1. Remove the 3/4 rack module top as described in
Replacement" on page
Step 2. Remove the A3 Analog Daughter Board as described in
Board Replacement" on page
Step 3. Disconnect the Mother-Analog Cable (item 1) from the A2 Analog Board.
Step 4. Disconnect the Analog Source Cable (item 2) from the A2 Analog Board.
Step 5. Remove the four 7 mm Standoff (item 3) from the A2 Analog Board.
Step 6. Remove the TORX T20 screw (item 4) from the PCA E4980 USB. Then,
Disconnect the ground cable from the PCA E4980 USB.
Step 7. Lift the A2 Analog Board and front panel toward upside.
Step 8. Remove the five 5/8 inch nuts and washers (item 5) fastening the BNC
connector.
Step 9. Remove the A2 Analog Board from Front Panel.
Step 10. Disconnect the two SMB cables (item 6) from the A2 Analog Board.
Replacement Procedure
Step 1. Reverse the order of the removal procedure.
When you fasten the A2 Analog Board with the five 5/8 inch nuts and
washers, set box torque wrench's torque to 17.4 in-lb.
When you fasten the A2 Analog Board with the four 7 mm Standoff, set
box torque wrench's torque to 8.7 in-lb.
Keysight E4980A/E4980AL Precision LCR Meter
Torque screwdriver, TORX T10
Torque screwdriver, TORX T20
Box torque wrench, 7 mm
Box torque wrench, 5/8 inch
Figure 6-5
and for this procedure.
73.
"3/4 Rack Module Top
77.
"A3 Analog Daughter
79