Measurement Configurations; State Analysis - HP 1652B Getting Started Manual

Logic
Hide thumbs Also See for 1652B:
Table of Contents

Advertisement

Minimum
Input Overdrive:
250 mV or 30% of the input amplitude,
whichever is
greater.
-
Maximum
Voltage:
2 40
volts peak.
Maximum
Power Available Through
Cables:
600 mA at SV per cable; 2 amp @
5V per HP 1652Bl1653B.
Measurement
Configurations
Analyzer
Configurations:
Analvzer 1
Timing
Off
State
Off
Timing
State
State
Off
Analvzer 2
Off
Timing
Off
State
State
Timing
State
Off
Channel Assignment:
Each group of 16 channels (a pod) can be assigned to
Analyzer 1, Analyzer 2, or remain unassigned. The HP 1652B contains 5 pods; the
HP 1653B contains
2
pods.
-
State Analysis
Memory
Data Acquisition:
1024 samples/channel.
Trace Specification
Clocks:
Five clocks (HP 1652B) or two clocks (HP 1653B) are available and can
be used by either one or two state analyzers at any time. Clock edges can be ORed
together and operate in single phase, two phase demultiplexing,
or two phase
mixed mode. The clock edge is selectable as positive, negative, or both edges for
each clock.
Clock Qualifier:
The high or low level of four ORed clocks (HP 1652B) or one
clock (HP1653B)
can be ANDed with the clock specification.
Setup time: 20 ns;
hold time: 5 ns.
Pattern Recognizers:
Each recognizer is the AND combination
of bit (0, 1, or X)
patterns in each label. Eight pattern recognizers are available when one state
analyzer is on. Four are available to each analyzer when two state analyzers are on.
Range Recognizers:
Recognizes data which is numerically
between or on two
specified patterns (ANDed combination
of zeros and/or ones). One range term is
available .and is assigned to the first state analyzer turned on. The maximum size is
32 bits and on a maximum of 2 pods.
General Information
1-8
HP 16528/1653B
Service
Manual