Gain Calibration
11. Disconnect all test equipment and all inputs to channel 1 and 2 of
the oscilloscope.
12. Select the Calibration choice field. When the pop-up appears,
select Gain, then select Start .
13. When the calibration is complete, the updated calibration status
screen will appear.
Trigger Calibration
14. Make sure all signals are disconnected from the channel 1 and 2
inputs of the oscilloscope.
15. Select the Calibration choice field. When the pop-up appears,
select Trigger level, then select Start.
16. When the calibration is complete, the updated calibration status
screen will appear.
Delay Calibration
17. Select the Calibration choice field. When the pop-up appears,
_
select Delay, then select Start.
18. Connect a BNC cable from the Probe Compensation output on
the rear panel, to the channel 1 input. The instrument will prompt
you when you need to switch to channel 2.
I
Note @
You have the option of using the 1O:l scope probe in place of the
recommended 1:l BNC cable. If you use the scope probe, you will
have to use the BNC-to-mini probe adapter supplied with the
instrument and set attenuation field in step 19 to 1O:l.
19. Set the attenuation field in the calibration screen to the
appropriate setting.
Installation, Maintenance
and Calibration
D-14
HP 16528/1653B
Front-Panel Reference