Oscilloscope Specifications; Horizontal; Trigger; Oscilloscope Operating Characteristics - HP 1652B Getting Started Manual

Logic
Hide thumbs Also See for 1652B:
Table of Contents

Advertisement

Oscilloscope
Specifications
Vertical
Horizontal
Time Interval
Measurement
Accuracy:
-
'(2%
X s/div + 0.01% X delta-t +
500 ps).
Trigger
Sensitivity:
10% of full screen.
The following specifications are the performance standards or limits against which
the oscilloscope in the HP 1652B/1653B is tested.
Bandwidth
(-3 dB):
dc to 100 MHz (single shot).
DC Gain Accuracy:
t 3%
of full scale.
DC Offset Accuracy:
t (2 mV + 2% of the channel offset + 2.5% of full scale).
Voltage Measurement
Accuracy
(DC):
(Gain accuracy + ADC resolution +
Offset accuracy).
-
Oscilloscope
The following operating characteristics are not specifications, but are typical
Operating
operating characteristics for the oscilloscope in the HP 1652B/1653B. These are
Characteristics
included as additional information
for the user.
VefiiCai
Transition
Time (10% to 90%): 5 3.5
ns.
(at BNC)
Number
of Channels:
2.
Vertical
Sensitivity
Range: 15
mV/div to 10 V/div (1:l probe).
Vertical
Sensitivity
Resolution:
Adjustable 2 digit resolution.
Maximum
Sample Rate:
400 MSamples/second.
Analog-to-Digital
Conversion:
6 bit real-time.
Analog-to-Digital
Resolution:
t 1.6%
of full scale.
Waveform
Record Length:
2048
points.
Input R:
1 MQ 2 1% or 50 Q t 1%.
Input
C: Approximately
7 pF.
Input Coupling:
dc.
HP 1652B/1653B
Service Manual
General Information
l-15