HP 1652B Getting Started Manual page 626

Logic
Hide thumbs Also See for 1652B:
Table of Contents

Advertisement

-
HP 16528/1653B
Performance
Tests
Service
Manual
3-39
6. Using the knob and SELECT key, set Input to CH 1, Probe to l:l, and
Impedance to 1 MOhm as in figure 3-46.
[Scope)-
Channel
(Autoscale
>
Input
[CH]
VfDiv
[lso.oJ
OffSEt
p5zzcir~
Probe
1-1
Impedance
[InOhm-J
Preset
pmq
s/Dlv
11o.001
Del a y
I
1 CH m'
:
:
:
: :
:
: :
: :
:
: :
:
:
,
,
1
1
I
1
I
L
I
I
Figure 3-46. Channel
Menu Configurktion
7. Press the TRACE/TRIG
key and set the Mode to Immediate
and Run mode
to Repetitive as in figure 3-47.
IScope
- Trigger
tlode
[ Immediate
]
Run mode
(Repetitive]
s/Dlv
[j
Delay
Armed
by
IRUn]
Figure 3-47. Trigger Menu Configuration