When the markers are set to Time,
you
can place the markers on the
waveforms at events of interest and the logic analyzer will tell you:
l
Time X to Trig(ger).
l
Time 0 to Trig(ger).
l
TimeXtoO.
To position the markers, move the cursor to the field of the marker you
wish to position and press SELECT. A pop-up will appear showing the
current time for that marker. Either rotate the KNOB or enter a
numeric value from the keypad to change the position of that marker.
Pressing SELECT when you are finished positions the marker and
closes the pop-up.
When the cursor is on either the X to Trig or 0 to Trig fields, you can
also enter a value directly from the keypad without pressing SELECT.
Timing Waveforms Menu
19-4
[i]
- lining
flarkers
1
Time
Accumulo te
IpiT-1
Tlme/Dlv
[500]
-70
ns
IC'OCK]
Figure
19-5.
Markers Time
The Time X to 0 field will change according to the position of the X
and 0 markers. If you place the cursor on the Time X to 0 field and
press SELECT, another pop-up will appear showing you all three
times: X to Trigger, 0 to Trigger, and Time X to 0.
[68000TItlNG]-
Timing
Waveforms
flarkers
-1
X to Trig L-4901
hcumul
ate
[ml
0 to Trig 11
Time/Div
1-1
Delay
I-1
[ tbrker
tloveamn t (=)I
4
Time
X 0 to to X Trigger
Trigger
to 0
-490
220
710 ns ns ns
Figure
19-6.
Time X to 0 Pop-up
If you rotate the KNOB while this pop-up is open, both X and 0
markers will move, but the relative placement between them will not
change.