Appendix I, Receiver Alerts, Alarm And Prompts - Dexcom G4 Professional User Manual

Continuous glucose monitoring system
Hide thumbs Also See for G4 Professional:
Table of Contents

Advertisement

CHAPTER 18: APPENDIX
APPENDIX I, RECEIVER ALERTS, ALARM AND
PROMPTS
The following tables describe the alarm, alerts and prompts and how the
receiver notifies you in both real-time and blinded modes.
Prompt - Shows on screen only. Silent, no vibrate or beep.
Alert - Notifies with vibrate and beep depending on the profile settings.
Alarm - Low 55 - Notifies with vibrate and beep. Cannot be changed.
Disabled in blinded mode.
Receiver Alerts, Alarm and Prompts
What will I see
on the receiver
screen?
In real-time mode:
In blinded mode:
No display
Prompt,
What does this
alert or
mean?
alarm?
Low
The most recent
glucose
sensor glucose
alarm
reading is at or
below 55 mg/dL.
Disabled
N/A
Dexcom G4 PLATINUM Professional System User's Guide
How will the
Will the
receiver notify
receiver
me?
re-notify
(vibrate and/or
me?
beep)
Vibrates 4
Yes, every
times and then
30 minutes
vibrates/beeps
after each
4 times every
confirmation
5 minutes until
until the
confirmed or the
blood
glucose value
glucose
goes above 55
value
mg/dL.
comes back
into range.
Disabled
Disabled
(continued on next page)
|