System Electromagnetic Immunity Specifications
Immunity Test
Conducted RF
IEC 61000- 4-6
(Receiver only)
Radiated RF
IEC 61000-4-3
210
|
Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System
IEC 60601
Receiver
Test Level
Compliance
Level
3 Vrms
3 Vrms
150 kHz to
80 MHz
10 V/m
3 V/m
80 MHz to
2.5 GHz
Electromagnetic
Environment
Guidance
Portable and mobile
RF communications
equipment should be
used no closer to any
part of the receiver,
including cables, than
the recommended
separation distance
calculated from the
equation applicable to
the frequency of the
transmitter.
Recommended
Separation Distance
d = 1.2 √P 150 kHz to
80 MHz
d = 1.2 √P 80 MHz to
800 MHz
d = 2.3 √P 800 MHz to
2.5 GHz
(continued on next page)