The total number of data pairs considered in this analysis was 9152. Of
these, eighty-two percent (82%) of the System readings fall within ±20
mg/dL (1.1 mmol/L) of the YSI blood glucose values (≤ 80 mg/dL or 4.4
mmol/L) and within ±20% of YSI blood glucose values (> 80 mg/dL or
4.4 mmol/L).
Table 1-A. System Agreement to YSI within CGM
Glucose Ranges
CGM
Glucose
Number
Range
of paired
mg/dL
CGM-YSI
(mmol/L)
Overall
40-60
(2.2-3.3)
61-80
(3.4-4.4)
81-180
(4.5-10.0)
181-300
(10.1-16.7)
301-350
(16.7-19.4)
351-400
(19.4-22.2)
NOTE: CGM readings are within 40-400 mg/dL (2.2-22.2 mmol/L),
inclusive.
Table 1-A is categorized within CGM glucose ranges. When you see
a CGM reading on your receiver, this table shows you how likely that
reading matches your blood glucose level (measured by YSI in the
study). For example, if your CGM reads 100 mg/dL (5.6 mmol/L), it is
Percent
within
15/15%
YSI
9152
71%
512
67%
781
73%
3853
67%
2784
72%
775
82%
447
74%
Dexcom G4 PLATINUM Professional System User's Guide
Percent
Percent
within
within
20/20%
30/30%
YSI
YSI
82%
92%
78%
88%
85%
94%
78%
91%
84%
93%
91%
97%
84%
91%
Percent
Percent
Greater
within
than
40/40%
40/40%
YSI
YSI
97%
3%
94%
6%
98%
2%
97%
3%
96%
4%
98%
2%
95%
5%
|
185
Need help?
Do you have a question about the G4 Professional and is the answer not in the manual?
Questions and answers