What To Do For Calibration Prompts; Calibration Error Troubleshooting - Dexcom G4 Professional User Manual

Continuous glucose monitoring system
Hide thumbs Also See for G4 Professional:
Table of Contents

Advertisement

13
13.2.2 WHAT TO DO FOR CALIBRATION
PROMPTS
1. When you see a calibration prompt, press the
the prompt.
2. Check the status area at the top of the
screen.
a. If the startup calibration symbol
shows, take 2 more blood glucose values
and enter them into the receiver.
b. If the additional startup calibration symbol
shows, take 1 more blood glucose
value and enter it into the receiver.
c. If the calibration needed symbol
glucose value and enter it into the receiver.
13.3 CALIBRATION ERROR
TROUBLESHOOTING
This screen means you recently entered
a calibration blood glucose value, and the
sensor is having trouble calibrating. If you
press the
SELECT
this symbol
If you see this screen, wait 15 minutes and
then enter 1 more calibration blood glucose
value. Wait 15 more minutes. If this error
screen still shows, enter 1 more blood glucose
value. Wait another 15 minutes. If this error screen still shows, the
sensor needs to be replaced.
This screen also means you recently entered a calibration blood glucose
value and the sensor is having trouble calibrating. If you press the
168
|
Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System
button to clear this screen,
shows in the status area.
SELECT
Prompt in status area
shows, take 1 more blood
Wait 15 minutes
calibration error screen
button to clear