glucose ranges above and below the paired CGM readings.
Table 3. Concurrence of CGM Readings and YSI Values
CGM
mg/dL
(mmol/L)
40-
< 40
60
(< 2.2)
(2.2-
3.3)
< 40
6%
48% 37%
(< 2.2)
40-60
4%
49% 36%
(2.2-3.3)
61-80
0%
22% 51%
(3.4-4.4)
81-120
0%
2%
(4.5-6.7)
121-160
0%
0%
(6.7-8.9)
161-200
0%
0%
(8.9-11.1)
201- 250
0%
0%
(11.1- 13.9)
251- 300
0%
0%
(13.9- 16.7)
301- 350
0%
0%
(16.7- 19.4)
351- 400
0%
0%
(19.4- 22.2)
> 400
0%
0%
(> 22.2)
Accuracy
Accuracy between matched pairs was also estimated by calculating the
percent difference between the System reading and the YSI value. For
example, if the YSI value is 100 mg/dL (5.6 mmol/L) and the System
YSI mg/dL (mmol/L)
Percent of matched pairs in each YSI glucose range
for each Sensor glucose range
61-
81-
121-
161-
80
120
160
200
(3.4-
(4.5-
(6.7-
(8.9-
4.4)
6.7)
8.9)
11.1)
7%
1%
0%
11%
1%
0%
24%
1%
0%
17%
66%
13%
1%
1%
25%
60%
13%
0%
2%
28%
53%
0%
0%
3%
21%
0%
0%
0%
4%
0%
0%
0%
0%
0%
0%
0%
0%
0%
0%
0%
0%
Dexcom G4 PLATINUM Professional System User's Guide
201-
251-
301-
351-
250
300
350
400
(11.1-
(13.9-
(16.7-
(19.4-
13.9)
16.7)
19.4)
22.2)
0%
0%
0%
0%
0%
0%
0%
0%
0%
0%
0%
0%
0%
0%
0%
0%
2%
0%
0%
0%
16%
2%
0%
0%
51%
21%
3%
1%
19%
49%
24%
3%
3%
28%
51%
16%
3%
10%
43%
38%
1%
6%
21%
57%
Number
of
> 400
Paired
(> 22.2)
CGM-
YSI
0%
155
0%
512
0%
781
0%
1706
0%
1492
0%
1240
0%
1181
0%
1018
1%
775
7%
447
15%
248
|
189

Advertisement