Regardless of which alert profile you choose, if you confirm the first
vibrate alert, you will not get a sound alert.
When you choose the profile setting this setting applies to all
alerts, alarm and prompts.
Within each profile setting, each alert has its own unique sound pattern,
tone and volume level. This lets you to easily identify each alert and
alarm and its meaning.
The fixed low alarm at 55 mg/dL cannot be turned off or adjusted. It is
disabled in blinded mode.
The soft, normal, attentive and HypoRepeat profiles have the following
alert sequence:
• The first alert is vibrate only.
• I f the alert is not confirmed in five minutes, the system vibrates and
beeps.
• I f the alert is not confirmed in five more minutes, the system vibrates
and beeps louder. This continues at the same volume every five
minutes until confirmed.
For the HypoRepeat profile in real-time mode only:
• I f the alert is confirmed and the sensor glucose readings continue to
be at or below 55 mg/dL the system repeats the alert sequence in 30
minutes.
Press the SELECT button to confirm any alert.

9.3.1 ALERT PROFILE OPTIONS

Vibrate profile: when you want to silence the receiver and be
alerted by vibration. The only exception to this is the fixed
low alarm at 55 mg/dL in real-time mode, which alerts you
as a vibration first, followed by beeps 5 minutes later if not
confirmed.
Dexcom G4 PLATINUM Professional System User's Guide
(continued on next page)
|
119