Receiver Electromagnetic Immunity Specifications
(continued from page before)
Immunity Test
IEC 61000-4-11
IEC 60601-1-11
Power
Frequency
(50/60 Hz)
Magnetic Field
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
Guidance and Manufacturer's Declaration –
Electromagnetic Immunity
The Dexcom G4 PLATINUM Professional System is intended for use
in the electromagnetic environment specified in the next table. The
customer or the user of the Dexcom G4 PLATINUM Professional
System should ensure that it is used in such an environment.
IEC 60601
Test Level
70% U
T
(30% dip in U
)
T
for 25 cycles
85% U
T
(15% dip in U
)
T
for 5 sec
< 5% U
T
(> 95% dip in U
)
T
for 5 sec
3 A/m
Dexcom G4 PLATINUM Professional System User's Guide
Receiver
Compliance
Level
70% U
T
(30% dip in U
)
T
for 25 cycles
85% U
T
(15% dip in U
)
T
for 5 sec
< 5% U
T
(> 95% dip in U
)
T
for 5 sec
3 A/m
Electromagnetic
Environment
Guidance
Power frequency
magnetic
fields should
be at levels
characteristic of
a typical location
in a typical
commercial
or hospital
environment.
|