14
The Dexcom G4 PLATINUM Professional Continuous Glucose
Monitoring System is safe for use on U.S. commercial airlines. The
Dexcom G4 PLATINUM Transmitter is an M-PED with emission levels
that meet RTCA/DO160, Section 21, Category M. Per FAA Advisory,
Circular #91-21, 1B, dated 8/25/06, any M-PED that meets this standard
in all modes may be used onboard the aircraft without any further
testing by the operator. This device can withstand exposure to common
electrostatic (ESD) and electromagnetic interference (EMI).
Guidance and Manufacturer's Declaration –
Electromagnetic Immunity
The transmitter (P/N 9438-01) is intended for use in the electromagnetic
environment specified in the next table. The customer or the user of the
transmitter should ensure that it is used in such an environment.
Transmitter Electromagnetic Immunity Specifications
Immunity Test
Electrostatic
Discharge
(ESD)
IEC 61000-4-2
Electrical Fast
Transient
/burst
204
|
Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System
IEC 60601 Test
Level
± 6 kV Contact
± 8 kV Air
± 2 kV for
power supply
lines
Transmitter
Electromagnetic
Compliance
Environment
Level
Guidance
± 6 kV
Floors should be
Contact
wood, concrete
or ceramic
± 8 kV Air
tile. If floors
are covered
with synthetic
material, the
relative humidity
should be at least
30%.
Not applicable
(continued on next page)