SECTION 4 — ON-SITE CALIBRATION PROCEDURES
Introduction and Purpose of Procedures
The following calibration procedures must be used whenever a CI-5000 Color Film Recorder, color
portrait camera, Matrox Illuminator board or strobe unit is replaced in the ID-3000 system, or
whenever a diagnostic procedure indicates the need for recalibrating any of these system
components.
Tools and Equipment Required:
Video Photometer (P/N 13149)
Oscilloscope - 40 MHz
Oscilloscope Probe (Tektronix 6131) (P/N CPS 212)
Multimeter/DVM
Final Assembly Alignment Gauge (P/N 13443)
Color Calibration Card (P/N 1B2195A)
Centronics Cable (P/N 1A5556A)
Electrostatic Discharge Warning
Sensitive electronic components in the CI-5000, computer and other parts of the system can easily be
damaged by small static discharges.
Always use a wrist strap and grounded anti-static mat
when removing equipment covers and servicing equipment.
4-3
Calibration Procedures