To Perform Hardware Self Test - Agilent Technologies InfiniiVision 3000 X-Series User Manual

Hide thumbs Also See for InfiniiVision 3000 X-Series:
Table of Contents

Advertisement

20
Utility Settings
Figure 50
4 Connect a BNC cable (40 inches maximum) from the TRIG OUT
5 Press the [Utility] key; then, press the Service softkey.
6 Begin the Self Cal by pressing the Start User Cal softkey.

To perform hardware self test

Pressing [Utility] > Service > Hardware Self Test performs a series of internal
procedures to verify that the oscilloscope is operating properly.
It is recommended you run Hardware Self Test:
• After experiencing abnormal operation.
• For additional information to better describe an oscilloscope failure.
• To verify proper operation after the oscilloscope has been repaired.
Successfully passing Hardware Self Test does not guarantee 100% of the
oscilloscope's functionality. Hardware Self Test is designed to provide an
80% confidence level that the oscilloscope is operating properly.
308
To Channel 1
To Channel 2
User Calibration cable for 4-channel oscilloscope
connector on the rear panel to the BNC barrel connector.
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
To Channel 3
Longer cable
to TRIG OUT
To Channel 4

Advertisement

Table of Contents
loading

Table of Contents