The Full Screen measurement interval variation measures the value on all
displayed data points.
The N Cycles measurement interval variation measures the value on an
integral number of periods of the displayed signal. If less than three edges
are present, the measurement shows "No edges".
The X cursors show what interval of the waveform is being measured.

Measurement Thresholds

Setting measurement thresholds defines the vertical levels where
measurements will be taken on an analog channel or math waveform.
Changing default thresholds may change measurement results
N O T E
The default lower, middle, and upper threshold values are 10%, 50%, and 90% of the value
between Top and Base. Changing these threshold definitions from the default values may
change the returned measurement results for Average, Delay, Duty Cycle, Fall Time,
Frequency, Overshoot, Period, Phase, Preshoot, Rise Time, +Width, and -Width.
1 From the Measurement Menu, press the Settings softkey; then, press the
2 Press the Source softkey to select the analog channel or math waveform
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
T
Thresholds softkey to set analog channel measurement thresholds.
You can also open the Measurement Threshold Menu by pressing
[Analyze] > Features and then selecting Measurement Thresholds.
source for which you want to change measurement thresholds.
Positive area
Negative area
Measurements
N cycles
14