20
Utility Settings
Figure 51
4 Connect a BNC cable (40 inches maximum) from the TRIG OUT
5 Press the [Utility] key; then, press the Service softkey.
6 Begin the Self Cal by pressing the Start User Cal softkey.
To perform hardware self test
Pressing [Utility] > Service > Hardware Self Test performs a series of internal
procedures to verify that the oscilloscope is operating properly.
It is recommended you run Hardware Self Test:
• After experiencing abnormal operation.
• For additional information to better describe an oscilloscope failure.
• To verify proper operation after the oscilloscope has been repaired.
342
For a 4- channel oscilloscope, connect BNC tees to the equal- length
cables as shown below. Then connect a BNC(f)- to- BNC(f) (barrel
connector) to the tee as shown below.
To Channel 1
To Channel 2
User Calibration cable for 4-channel oscilloscope
connector on the rear panel to the BNC barrel connector.
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
To Channel 3
Longer cable
to TRIG OUT
To Channel 4