4 Press the [Single] key (in the Run Control group on the front panel) to
5 Press the Scroll Lister softkey and use the Entry knob to scroll through
6 Press the Zoom to Selection softkey (or push the Entry knob) to center
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
select All, the decode information for different buses is interleaved in
time.)
Before you can select a row or navigate through the Lister data,
oscilloscope acquisitions must be stopped.
stop the acquisition.
Pressing [Single] instead of [Stop] fills the maximum memory depth.
When zoomed out and viewing a large number of packets, the Lister
may not be able to display information for all packets. However, when
you press the [Single] key the Lister will contain all on- screen serial
decode information.
the data.
Time tags in the Time column indicate the event time relative to the
trigger point. The time tags of events that are shown in the waveform
display area are displayed with a dark background.
the waveform display at the time associated with the selected Lister
row and automatically set the horizontal scale setting.
7
Serial Decode
127