10
Triggers
7 Press the Auto Setup softkey to automatically set up the oscilloscope for
8 Press the Mode softkey to select the portion of the video signal that you
166
The Generic selection lets you trigger on custom bi- level and tri- level
sync video standards. See
page 168.
the selected Source and Standard:
Source channel vertical scaling is set to 140 mV/div.
Source channel offset is set to 245 mV.
Source channel is turned on.
Trigger type is set to Video.
Video trigger mode is set to All Lines (but left unchanged if Standard is
Generic).
Display Grid type is set to IRE (when Standard is NTSC) or mV (see
select the grid type"
Horizontal time/division is set to 10 µs/div for NTSC/PAL/SECAM
standards or 4 µs/div for EDTV or HDTV standards (unchanged for
Generic).
Horizontal delay is set so that trigger is at first horizontal division
from the left (unchanged for Generic).
You can also press [Analyze]> Features and then select Video to quickly
access the video triggering automatic set up and display options.
would like to trigger on.
The Video trigger modes available are:
Field1 and Field2 — Trigger on the rising edge of the first serration
pulse of field 1 or field 2 (interlaced standards only).
All Fields — Trigger on the rising edge of the first pulse in the vertical
sync interval.
All Lines — Trigger on all horizontal sync pulses.
Line — Trigger on the selected line number (EDTV and HDTV
standards only).
Line: Field1 and Line:Field2 — Trigger on the selected line # in field 1 or
field 2 (interlaced standards only).
Line: Alternate — Alternately trigger on the selected line # in field 1
and field 2 (NTSC, PAL, PAL- M, and SECAM only).
"To set up Generic video triggers"
on page 134).
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide