Triggers
7 Press the Auto Setup softkey to automatically set up the oscilloscope for
8 Press the Mode softkey to select the portion of the video signal that you
166
The Generic selection lets you trigger on custom bi- level and tri- level
sync video standards. See
page 168.
the selected Source and Standard:
•
Source channel vertical scaling is set to 140 mV/div.
•
Source channel offset is set to 245 mV.
•
Source channel is turned on.
Trigger type is set to Video.
•
Video trigger mode is set to All Lines (but left unchanged if Standard is
•
Generic).
Display Grid type is set to IRE (when Standard is NTSC) or mV (see
•
select the grid type"
•
Horizontal time/division is set to 10 µs/div for NTSC/PAL/SECAM
standards or 4 µs/div for EDTV or HDTV standards (unchanged for
Generic).
•
Horizontal delay is set so that trigger is at first horizontal division
from the left (unchanged for Generic).
You can also press [Analyze]> Features and then select Video to quickly
access the video triggering automatic set up and display options.
would like to trigger on.
The Video trigger modes available are:
Field1 and Field2 — Trigger on the rising edge of the first serration
•
pulse of field 1 or field 2 (interlaced standards only).
All Fields — Trigger on the rising edge of the first pulse in the vertical
•
sync interval.
All Lines — Trigger on all horizontal sync pulses.
•
Line — Trigger on the selected line number (EDTV and HDTV
•
standards only).
Line: Field1 and Line:Field2 — Trigger on the selected line # in field 1 or
•
field 2 (interlaced standards only).
Line: Alternate — Alternately trigger on the selected line # in field 1
•
and field 2 (NTSC, PAL, PAL- M, and SECAM only).
"To set up Generic video triggers"
on page 134).
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
on
"To