Agilent Technologies InfiniiVision 3000 X-Series User Manual page 344

Hide thumbs Also See for InfiniiVision 3000 X-Series:
Table of Contents

Advertisement

22
Reference
Number of
Waveform
Buffers
Points
Count
X Display Range
X Display Origin
X Increment
X Origin
X Units
Y Units
Date
Time
344
A 32-bit integer that is the number of waveform buffers required to read the
data.
A 32-bit integer that is the number of waveform points in the data.
A 32-bit integer that is the number of hits at each time bucket in the waveform
record when the waveform was created using an acquisition mode like
averaging. For example, when averaging, a count of four would mean every
waveform data point in the waveform record has been averaged at least four
times. The default value is 0.
A 32-bit float that is the X-axis duration of the waveform that is displayed. For
time domain waveforms, it is the duration of time across the display. If the
value is zero then no data has been acquired.
A 64-bit double that is the X-axis value at the left edge of the display. For time
domain waveforms, it is the time at the start of the display. This value is
treated as a double precision 64-bit floating point number. If the value is zero
then no data has been acquired.
A 64-bit double that is the duration between data points on the X axis. For
time domain waveforms, this is the time between points. If the value is zero
then no data has been acquired.
A 64-bit double that is the X-axis value of the first data point in the data
record. For time domain waveforms, it is the time of the first point. This value
is treated as a double precision 64-bit floating point number. If the value is
zero then no data has been acquired.
A 32-bit integer that identifies the unit of measure for X values in the acquired
data:
• 0 = Unknown.
• 1 = Volts.
• 2 = Seconds.
• 3 = Constant.
• 4 = Amps.
• 5 = dB.
• 6 = Hz.
A 32-bit integer that identifies the unit of measure for Y values in the acquired
data. The possible values are listed above under X Units.
A 16-byte character array, left blank in InfiniiVision oscilloscopes.
A 16-byte character array, left blank in the InfiniiVision oscilloscopes.
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide

Advertisement

Table of Contents
loading

Table of Contents