28
UART/RS232 Triggering and Serial Decode
4 Optional: Press the Framing softkey and select a value. In the decode
5 If the decode line does not appear on the display, press the [Serial] key
6 If the oscilloscope is stopped, press the [Run/Stop] key to acquire and
If the setup does not produce a stable trigger, the UART/RS232 signals may be slow
N O T E
enough that the oscilloscope is AutoTriggering. Press the [Mode/Coupling] key, then
press the Mode softkey to set the trigger mode from Auto to Normal.
You can use the horizontal Zoom window for easier navigation of the
acquired data.
See Also
428
When #Bits is set to 9 in the UART/RS232 Bus Configuration Menu,
the 9th (alert) bit is displayed directly to the left of the ASCII value
(which is derived from the lower 8 bits).
display, the chosen value will be displayed in light blue. However, if a
parity error occurs the data will be displayed in red.
to turn it on.
decode data.
"Interpreting UART/RS232 Decode"
"UART/RS232 Totalizer"
"Interpreting UART/RS232 Lister Data"
"Searching for UART/RS232 Data in the Lister"
on page 429
on page 430
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
on page 431
on page 431