Building A Mask File - Agilent Technologies InfiniiVision 3000 X-Series User Manual

Hide thumbs Also See for InfiniiVision 3000 X-Series:
Table of Contents

Advertisement

15
Mask Testing
The mask scaling controls how the normalized vectors are interpreted.
This in turn controls how the mask is drawn on the display. The remote
programming commands that control mask scaling are:
:MTES:SCAL:BIND 0
:MTES:SCAL:X1 -400.000E-06
:MTES:SCAL:XDEL +800.000E-06
:MTES:SCAL:Y1 +359.000E-03
:MTES:SCAL:Y2 +2.35900E+00

Building a Mask File

The following mask uses all eight mask regions. The most difficult part of
creating a mask file is normalizing the X and Y values from the time and
voltage values. This example shows an easy way to convert voltage and
time to normalized X and Y values in the mask file.
The following mask file produced the mask shown above:
MASK_FILE_548XX
"All Regions"
/* Region Number */ 1
/* Number of vertices */ 4
252
-12.50,
MAX
-10.00,
1.750
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide

Advertisement

Table of Contents
loading

Table of Contents