6 Press the qualifier softkey (< > ><) to select the time qualifier.
7 Select the qualifier time set softkey (< or >), then rotate the Entry
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
When triggering on a positive pulse, the trigger will occur on the high
to low transition of the pulse if the qualifying condition is true. When
triggering on a negative pulse, the trigger will occur on the low to high
transition of the pulse if the qualifying condition is true.
The Qualifier softkey can set the oscilloscope to trigger on a pulse
width that is:
Less than a time value (<).
For example, for a positive pulse, if you set t<10 ns:
10 ns
Greater than a time value (>).
For example, for a positive pulse, if you set t>10 ns:
10 ns
Within a range of time values (><).
For example, for a positive pulse, if you set t>10 ns and t<15 ns:
10 ns
knob to set the pulse width qualifier time.
The qualifiers can be set as follows:
2 ns to 10 s for > or < qualifier (5 ns to 10 s for 350 MHz
bandwidth models).
10 ns to 10 s for >< qualifier, with minimum difference of 5 ns
between upper and lower settings.
10 ns
15 ns
Triggers
10 ns
Trigger
Trigger
Trigger
12 ns
10
151