14
Measurements
See Also
"To isolate a pulse for Top measurement"
on page 223
To isolate a pulse for Top measurement
The following figure shows how to use Zoom mode to isolate a pulse for a
Top measurement.
You may need to change the measurement window setting so that the
measurement is made in the lower, Zoom window. See
"Measurement
Window with Zoom Display"
on page 239.
Figure 46
Isolating area for Top measurement

Base

The Base of a waveform is the mode (most common value) of the lower
part of the waveform, or if the mode is not well defined, the base is the
same as Minimum. The Y cursor shows the value being measured.
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
223