1 Press [Utility] > Options > Rear Panel.
2 In the Rear Panel Menu, press Trig Out; then, turn the Entry knob to
The TRIG OUT connector also provides the User Cal signal. See
perform user calibration"

Performing Service Tasks

The Service Menu (under [Utility] > Service) lets you perform service- related
tasks:
For other information related to oscilloscope maintenance and service, see:
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
select from:
Triggers— Each time the oscilloscope triggers, a rising edge occurs on
TRIG OUT. The rising edge is delayed 30 ns from the oscilloscope's
trigger point. The output level is 0- 5 V into an open circuit, and
0- 2.5 V into 50
Mask— The pass/fail status is evaluated periodically. When the
evaluation of the testing period results in a failure, the trigger output
pulses high (+5 V). Otherwise, the trigger output remains at low
(0 V). See
Chapter
Waveform Generator Sync Pulse— All of the waveform generator output
functions (except DC, Noise, and Cardiac) have an associated Sync
signal:
The Sync signal is a TTL positive pulse that occurs when the
waveform rises above zero volts (or the DC offset value).
See
Chapter
17, "Waveform Generator," starting on page 259.
"To perform user calibration"
"To perform hardware self test"
"To perform front panel self test"
"To display oscilloscope information"
"To display the user calibration status"
. See
Chapter
10, "Triggers," starting on page 143.
15, "Mask Testing," starting on page 243.
on page 306.
on page 306
on page 308
on page 309
on page 309
Utility Settings
on page 309
20
"To
305