MIL-STD-1553/ARINC 429 Triggering and Serial Decode
3 Press the Word Format softkey and specify the word decode format:
4 If the decode line does not appear on the display, press the [Serial] key
5 If the oscilloscope is stopped, press the [Run/Stop] key to acquire and
If the setup does not produce a stable trigger, the ARINC 429 signal may be slow enough
N O T E
that the oscilloscope is AutoTriggering. Press the [Mode/Coupling] key, then press the
Mode softkey to set the trigger mode from Auto to Normal.
You can use the horizontal Zoom window for easier navigation of the
decoded data.
See Also
418
The base setting is used for data display in both the decode line and in
the Lister.
Label values are always displayed in octal, and SSM and SDI values are
always displayed in binary.
Label/SDI/Data/SSM:
Label - 8 bits.
SDI - 2 bits.
Data - 19 bits.
SSM - 2 bits.
Label/Data/SSM:
Label - 8 bits.
Data - 21 bits.
SSM - 2 bits.
Label/Data:
Label - 8 bits.
Data - 23 bits.
to turn it on.
decode data.
"Interpreting ARINC 429 Decode"
"ARINC 429 Totalizer"
"Interpreting ARINC 429 Lister Data"
"Searching for ARINC 429 Data in the Lister"
on page 419
on page 420
on page 421
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
on page 421