Download Print this page

Impulse Dynamics OPTIMIZER Smart Mini Instructions For Use Manual page 74

Hide thumbs Also See for OPTIMIZER Smart Mini:

Advertisement

3.21.3 Charger Failure Alert
The Charger Failure alert is triggered when the Guardio/Vesta Charger detects an
internal failure.
To set the Charger Failure alert:
1. If necessary, initiate a link with OPTIMIZER Smart Mini IPG
a. Place (or replace, if necessary) the Intelio Programming Wand over the
OPTIMIZER Smart Mini IPG implant site
b. Tap the Start OPTIlink button on the OPTIlink Session Pane
2. Tap the Preferences button on the Menu Bar
3. Select the Patient Alerts tab
4. On the Patient Alerts panel, tap the Charger Failure toggle switch to enable
(ON) or disable (OFF) the alert
5. Tap the blinking Program button on the Programming Buttons Pane to load the
OPTIMIZER Smart Mini IPG with the new parameter setting
3.21.4 Battery Low Alert
The Battery Low alert is triggered whenever the OPTIMIZER Smart Mini IPG's battery
voltage drops to 3.5 V.
To set the Battery Low alert:
1. If necessary, initiate a link with OPTIMIZER Smart Mini IPG
a. Place (or replace, if necessary) the Intelio Programming Wand over the
OPTIMIZER Smart Mini IPG implant site
b. Tap the Start OPTIlink button on the OPTIlink Session Pane
2. Tap the Preferences button on the Menu Bar
3. Select the Patient Alerts tab
4. On the Patient Alerts panel, tap the Battery Low toggle switch to enable (ON)
or disable (OFF) the alert
5. Tap the blinking Program button on the Programming Buttons Pane to load the
OPTIMIZER Smart Mini IPG with the new parameter setting
3.21.5 Battery Recharge Reminder Alert
The Guardio/Vesta Charger keeps a record of OPTIMIZER Smart Mini IPG charging
events. This record is used to determine the number of days since it last charged the
OPTIMIZER Smart Mini IPG.
The Battery Recharge Reminder alert is triggered if the number of days since the
Guardio/Vesta Charger lasted charged the OPTIMIZER Smart Mini IPG exceeds the
Battery Recharge Reminder days parameter value set by the Optimizer Smart Mini
Programmer application.
To set the Battery Recharge Reminder alert and set its days value:
1. If necessary, initiate a link with OPTIMIZER Smart Mini IPG
a. Place (or replace, if necessary) the Intelio Programming Wand over the
OPTIMIZER Smart Mini IPG implant site
b. Tap the Start OPTIlink button on the OPTIlink Session Pane
2. Tap the Preferences button on the Menu Bar
3. Select the Patient Alerts tab
4. On the Patient Alerts panel, tap the Battery Recharge Reminder toggle switch
to enable (ON) or disable (OFF) the alert
64

Advertisement

loading
Need help?

Need help?

Do you have a question about the OPTIMIZER Smart Mini and is the answer not in the manual?

Questions and answers

Related Products for Impulse Dynamics OPTIMIZER Smart Mini

This manual is also suitable for:

Optimizer smart systems