7.1.11 Statistics Menu
The Statistics menu contains the following tabs, each with their own panel:
On - General: displays the number of sensed events, periods, and trains during
scheduled CCM delivery.
On - Inhibition: displays the number of sensed events that inhibited CCM during
scheduled CCM delivery.
Off - General: displays the number of sensed events and periods when CCM is
not scheduled for delivery.
Off - Inhibition: displays the number of sensed events that would have inhibited
CCM when CCM is not scheduled for delivery.
Last Session: displays the number of ventricular events and trains delivered
during the last scheduled CCM delivery session, the percentage of CCM
delivered during the last scheduled CCM delivery session, and the number of
battery discharge episodes.
The panel of each tab can be viewed in Numeric Mode or Graph Mode
Figure 98: Statistics On - General Panel (Numeric View)
Figure 97: Settings Tab