7.12.2 Turning Markers Off
To turn off Marker Mode:
1. Tap the Marker button the Tool Bar, or
2. Remove the Legacy Programming Wand from the implant zone. If the Legacy
Programming Wand cannot communicate with the OPTIMIZER Smart IPG, the
Marker Mode is automatically terminated.
In all cases, the Programmer reports "Marker mode end OK".
Whenever a communication command between the OPTIMIZER Smart IPG and the
Intelio Programmer takes place (programming, reading statistics, measuring impedances,
etc.), the Marker Mode automatically ends and is re-established when the communication
has finished.
7.12.3 Sensed Events
Sensed events are represented by colored line markers below the baseline, which use
the following color conventions:
Magenta: Atrial sensed event
Green: Ventricular sensed event
Black: Local Sense event
7.12.4 CCM Therapy Delivery
CCM therapy delivery is represented by a marker in the shape of a blue rectangle whose
width is in proportion to the duration of the CCM pulse train.
7.12.5 CCM Inhibit Conditions
Labels for the following events are displayed:
PVC: two adjacent ventricular sensed events without an intervening atrial sensed
event.
AT: Atrial rate above the atrial tachycardia rate limit. Any atrial rate exceeding a
certain threshold is considered atrial tachycardia. Using the Optimizer Smart
Programmer application, the atrial tachycardia threshold rate can be set to one of
51 possible values between 62 bpm and 179 bpm. CCM therapy delivery is
always inhibited when atrial tachycardia is detected.
A Noise: Noise detected on the atrial sense channel.
V Noise: Noise detected on the ventricular sense channel.
Long AV: Sensed AV interval exceeds "Long AV" limit.
Short AV: Sensed AV interval shorter than "Short AV" limit.
7.12.6 CCM Therapy Delivery State
CCM therapy delivery state is indicated by labels as well as by the background color of
the Marker window as follows:
Inactive: this label appears on the Marker window when CCM delivery first
becomes not Active. The background color of the Marker window is white in this
case.
Active-On: this label appears on the Marker window when CCM delivery first
becomes Active and On. The background color of the Marker window is green
in this case.
Active-Off: this label appears on the Marker window when CCM delivery first
becomes Active but Off. The background color of the Marker window is gray in
this case.