Figure 44: OPTIset: IEGM SENSITIVITIES Window
When the OPTIset: CCM ALGORITHM TIMING window appears, the analysis will begin
automatically. When the analysis is done, the buttons at the bottom right of the screen will
become enabled (turn blue).
10. Tap the Propose button
If the analysis was successful, a green checkmark will be shown along with a message
indicating that the OPTIset proposal of CCM algorithm timing is complete. If any proposed
parameter value is different from the currently programmed value, it will be shown in blue.
Note: The proposed parameter values can be modified by tapping on the specific parameter
and choosing a new value.
Note: If OPTIset cannot successfully find a set of proposed values, tap the Repeat button to
repeat the analysis. If OPTIset still cannot successfully find a set of proposed values after the
second attempt, tap the Accept & Exit Optiset button to close the OPTIset application and
set the CCM Timing parameters on the CCM Timing panel manually.
11. Tap the Accept & Continue to Optiset Amplitude button.