LS Absence: The number of ventricular events detected without
corresponding Local Sense events during scheduled CCM
delivery.
VT: The number of ventricular events exceeding the tracking rate
during scheduled CCM delivery (OVO-LS-CCM mode only).
Charger: The number of normal event cycles where CCM
delivery was inhibited during scheduled CCM delivery due to the
IPG being subjected to a charging session.
Noise Episodes
o
A Noise: The number of atrial noise episodes detected during
scheduled CCM delivery (ODO-LS-CCM mode only).
V Noise: The number of ventricular noise episodes detected
during scheduled CCM delivery.
•
Off - General: Events occurring when the OPTIMIZER Smart Mini IPG is
not scheduled to deliver CCM therapy (Off Time). Displays the same list
of CCM statistical counters as detailed under On - General (except for
Trains Delivered).
•
Off - Inhibition: Inhibitions occurring when the OPTIMIZER Smart Mini
IPG is not scheduled to deliver CCM therapy (Off Time). Displays the
same list of CCM statistical counters as detailed under On - Inhibition.
•
Other
Last Session: The period between the previous day's CCM Therapy
o
Start Time and End Time.
Sensed V: The number of ventricular events sensed during the
previous day's CCM scheduled delivery sessions.
Delivered Trains: The number of CCM trains delivered during
the previous day's CCM scheduled delivery sessions.
CCM%: CCM delivery percentage during the previous day's
CCM scheduled delivery sessions.
Lead Displacement Alert: Indicates whether the Maximum lead
impedance change alert was triggered as a result of the previous
day's impedance measurement.
CCM % Low Alert: Indicates whether the Minimum CCM% alert
was triggered as a result of the previous day's CCM scheduled
delivery sessions.
General
o
Lead Displacement: The number of lead displacement
detections.
Battery Discharge Episodes: The number of times the device
reverted to OOO mode due to the OPTIMIZER Smart Mini IPG's
battery voltage dropping to 3.5 V.
3.15.1.3 CCM Statistics Buttons
The CCM Statistics panel has the following buttons:
•
Read: Reads the latest CCM statistics from OPTIMIZER Smart Mini IPG.
•
Reset: Resets the CCM statistics stored in OPTIMIZER Smart Mini IPG.
48