Figure 47: CCM Statistics
3.15.1.2 CCM Statistics Tabs
The following are the list of tabs for the CCM Statistics and the statistics that are
shown under each tab:
•
On - General: Events occurring when the OPTIMIZER Smart Mini IPG is
scheduled to deliver CCM therapy (On Time).
Events
o
Atrial: The number of atrial events sensed during scheduled
CCM delivery (ODO-LS-CCM mode only).
Ventricular: The number of ventricular (RV) events sensed
during scheduled CCM delivery.
LS in Alert: The number of local sense (LS) events sensed
during scheduled CCM delivery.
Periods
o
Normal: The number of normal event cycles (i.e., detected
ventricular and LS in Alert events) during scheduled CCM
delivery.
Inhibited: The number of inhibited event cycles during
scheduled CCM delivery.
Post-Inhibited: The number of post-inhibited event cycles during
scheduled CCM delivery.
Trains Delivered
o
Total: The total number of CCM trains delivered during
scheduled CCM delivery.
•
On - Inhibition: Inhibitions occurring when the OPTIMIZER Smart Mini
IPG is scheduled to deliver CCM therapy (On Time).
Causes
o
AT: The number of atrial events exceeding the tracking rate
during scheduled CCM delivery (ODO-LS-CCM mode only).
PVC: The number of PVCs detected during scheduled CCM
delivery (ODO-LS-CCM mode only).
Long AV: The number of times Long AV condition detected
during scheduled CCM delivery (ODO-LS-CCM mode only).
Short AV: The number of times Short AV condition detected
during scheduled CCM delivery (ODO-LS-CCM mode only).
LS Out of Window: The number of Local Sense events outside
the LS Alert Window during scheduled CCM delivery.