GUIDELINES AND MANUFACTURER'S DECLARATION – ELECTROMAGNETIC EMMISSIONS OF
THE INTELIO PROGRAMMER WITH LEGACY PROGRAMMING WAND PURSUANT TO:
ETSI EN 302 195 V2.1.1 – Short Range Devices (SRD); Ultra Low Power Active Medical Implants
(ULP-AMI) and accessories (ULP-AMI-P) operating in the frequency range 9 kHz to 315 kHz
Harmonised Standard covering the essential requirements of article 3.2 of the Directive
2014/53/EU
The Intelio Programmer with Legacy Programming Wand, part of the OPTIMIZER Smart Mini System,
is intended for use in an electromagnetic environment as specified below. The customer or user of the
Intelio Programmer must ensure that it is used within the specified environment.
Emissions Test
Radiated Field Strength
Modulation Bandwidth
Transmitter Spurious Emissions
(9 kHz to 30 MHz)
Duty Cycle
Receiver Blocking
Receiver Spurious Emissions
(9 kHz to 30 MHz)
GUIDELINES AND MANUFACTURER'S DECLARATION – ELECTROMAGNETIC EMMISSIONS OF
THE INTELIO PROGRAMMER WITH LEGACY PROGRAMMING WAND PURSUANT TO:
ETSI EN 301 489-1 V2.2.3 – ElectroMagnetic Compatibility (EMC) standard for radio equipment
and services; Part 1: Common technical requirements; Harmonised Standard for
ElectroMagnetic Compatibility
EN 301 489-31 – ElectroMagnetic Compatibility (EMC) standard for radio equipment and
services; Part 31: Specific conditions for equipment in the 9 kHz to 315 kHz band for Ultra Low
Power Active Medical Implants (ULP-AMI) and related peripheral devices (ULP-AMI-P);
Harmonised Standard covering the essential requirements of article 3.1(b) of the Directive
2014/53/EU
The Intelio Programmer with Legacy Programming Wand, part of the OPTIMIZER Smart Mini System,
is intended for use in an electromagnetic environment as specified below. The customer or user of the
Intelio Programmer must ensure that it is used within the specified environment.
There may be potential difficulties in ensuring electromagnetic compatibility in other environments, due
to conducted as well as radiated disturbances.
Emissions Test
Radiated Emissions
Conducted Emissions
Harmonic Current
Emissions
Voltage Fluctuations
ETSI EN 302 195
Compliance
Complies with clause 4.2.1
Complies with clause 4.2.2
Complies with clause 4.2.3
Complies with clause 4.2.4
Complies with clause 4.3.2
Complies with clause 4.3.3
ETSI EN 301 489-1 and EN 301 489-31
Basic Standard
EN 55032
EN 55032
IEC 61000-3-2
IEC 6100-3-3
134
environment – guidelines
The Intelio Programmer with
Legacy Programming Wand
must emit electromagnetic
energy in order to perform its
intended function. Nearby
electronic equipment may be
affected.
Compliance
environment – guidelines
Pass
The Intelio Programmer with
Legacy Programming Wand
Pass
must emit electromagnetic
Pass
energy in order to perform its
intended function. Nearby
electronic equipment may be
Pass
affected.
Electromagnetic
Electromagnetic
Need help?
Do you have a question about the OPTIMIZER Smart Mini and is the answer not in the manual?
Questions and answers