7.16.1 Loading Statistics
To obtain the Statistics from the OPTIMIZER Smart IPG:
1. Place (or replace, if necessary) the Legacy Programming Wand over the
OPTIMIZER Smart IPG implant site.
2. Tap the Statistics button on the Menu Bar.
3. Tap the Read button at the bottom of the Statistics panel.
If the loading is successful, the Programmer will display the message "Read Statistics
OK".
However, if the Legacy Programming Wand is not well positioned over the implant site,
the transfer of data may fail. If a failure in communication occurs, the Programmer will
display the message "Read Statistics Error" along with the options Retry and Cancel.
Retry can be performed by repositioning the Legacy Programming Wand and selecting
the Retry button.
When the OPTIMIZER Smart IPG is in Marker Mode and within communications range
of the Legacy Programming Wand, the contents of the Statistics counters change
dynamically, showing the current number of occurrences of each event type.
7.16.2 Statistics Tabs
The following are the list of tabs for the Statistics and the statistics that are shown under
each tab:
•
On – General: Events occurring when the OPTIMIZER Smart IPG is scheduled
to deliver CCM therapy (On Time).
Events
o
Atrial: number of atrial events sensed during scheduled CCM delivery
Ventricular: number of ventricular events sensed during scheduled CCM
delivery
LS in Alert: number of local sense events sensed during scheduled
CCM delivery
Periods
o
Normal: number of normal periods during scheduled CCM delivery
Inhibited: number of inhibited periods during scheduled CCM delivery
Post-Inhibited: number of post-inhibited periods during scheduled CCM
delivery
Trains Delivered
o
During Onset: number of CCM trains delivered during the onset of
scheduled CCM delivery
Total: total number of CCM trains delivered during scheduled CCM
delivery
•
On – Inhibition: Inhibitions occurring when the OPTIMIZER Smart IPG is
scheduled to deliver CCM therapy (On Time).
Causes
o
AT: number of atrial tachycardia beats detected during scheduled CCM
delivery
PVC: number of PVCs detected during scheduled CCM delivery
Long AV: number of times Long AV condition detected during scheduled
CCM delivery