Figure 45: OPTIset: CCM ALGORITHM TIMING Window
When the OPTIset: CCM AMPLITUDE window appears, the analysis will begin automatically
if at least one CCM Delivery Channel has been enabled.
Note: If no CCM Delivery channels have been enabled, enable one CCM Delivery channel,
wait for OPTIset to complete its analysis, and then enable the second CCM Delivery channel.
Note: As long as at least one CCM Delivery channel is enabled, enabling or disabling a CCM
Delivery channel will cause OPTIset to automatically begin the analysis of the CCM
amplitude again.
When the analysis is done, the buttons at the bottom right of the screen will become enabled
(turn blue).
12. Tap the Propose button
If the analysis was successful, a green checkmark will be shown along with a message
indicating that the OPTIset proposal of CCM amplitude is complete. If the proposed CCM
Amplitude parameter value is different from the currently programmed value, it will be shown
in blue.
Note: The proposed CCM Amplitude parameter value can be modified by tapping on the
parameter and choosing a new value.
Note: If OPTIset cannot successfully find a proposed CCM Amplitude parameter, tap the
Repeat button to repeat the analysis. If OPTIset still cannot successfully find a proposed
CCM Amplitude parameter after the second attempt, tap the Accept & Continue button to
close the OPTIset application and set the CCM Amplitude parameter on the CCM Train
panel manually.
13. Tap the Accept & Continue button on the OPTIset: CCM AMPLITUDE window.