Required Test Equipment - Agilent Technologies N9039A Service Manual

Rf preselector
Hide thumbs Also See for N9039A:
Table of Contents

Advertisement

Performance Verification and Adjustment Software

Required Test Equipment

Required Test Equipment

The following table identifies the equipment required for verifying the
performance of the N9039A RF Preselector as a stand alone instrument. Only the
recommended and alternate equipment is compatible with the performance
verification testing. Some tests can use various models of a particular equipment
type. The "Recommended Agilent Model" is the preferred equipment. However,
the "Alternative Agilent Model" is an acceptable substitute.
Table 10-1
Required Test Equipment for N9039A RF Preselector
Performance Tests
Instrument
Signal Sources
DC Pulse Generator
EMC System Source
Microwave Source 1
244
Critical Specifications
Frequency: 9 kHz to 1.0 GHz
Frequency: 100 kHz to 1.0 GHz
Spectral Purity
SSB Phase Noise @ 1 GHz:
−145 dBc at 100 kHz offset
−158 dBc at 1 MHz offset
−160 dBc at 6 MHz offset
−160 dBc at 10 MHz offset
Harmonics:
−30 dBc @ ≤ +10 dBm output
Frequency: 10 MHz to 1.0 GHz
Frequency Resolution: 1 Hz
Harmonic level: < −40 dBc
Amplitude range: −20 dBm to +13
Amplitude resolution: 0.02
Harmonic level:
< 2 GHz, < -30 dBc
VSWR: < 1 GHz: 1.6:1
Recommended
Alternative
Agilent Model
Agilent Model
Number
Number
Schwarzbeck
IGUU2916
E8257D
E4438C
N5181A
b
83630A/B,
PSG
83640A/B,
83650A/B
(Option 001,008)
E8241A, E8244A,
E8251A, E8254A,
E8247C, E8257C
(Opt. H31 or 540,
1E1, 1EA)
E8247C, E8257C
(Opt. 1E1, 1EA,
520)
E8257D
(Options 1EA, 567)
E8267C
(Option 520)
E8267D
(Option 544 or 532
or 520)
a
Use
P
P
P
Chapter 10