Assembly Replacement Procedures
Input and Output Connectors

Replacement

1. Refer to
Figure
7-27. Place the connector assembly into position in the chassis.
Replace the two screws (1) to attach the connector assembly to the chassis.
Torque to 9 inch-pounds.
2. Reconnect the semi-rigid cable to the connector. Torque to 10 inch-pounds.
3. Replace the Front Frame Assembly. Refer to the
replacement procedure.
4. Replace the instrument outer case. Refer to the
replacement procedure.
204
Front Frame Assembly
Instrument Outer Case
Chapter 7