Midplane Board Assembly

Removal

When the A7 Midplane board assembly is replaced all installed instrument license
keys will be lost. So, if at all possible, backup the installed license keys before
replacing the assembly. If the license keys are not backed up prior to replacing the
board they will need to be retrieved from the Agilent Software Licensing website
once the board is replaced and programmed with the correct instrument model and
serial numbers.
For instructions on backing up the instrument license keys see
Backup and Restoring"
1. Remove the instrument outer case. Refer to the
procedure.
2. Remove the processor assembly. Refer to the
procedure.
3. Remove the rear panel. Refer to the
4. Remove the top brace. Refer to the
5. Remove the power supply assembly. Refer to the
removal procedure.
Chapter 7
in
Chapter 8 , "Post-Repair
Rear Panel
Top Brace
Assembly Replacement Procedures

Midplane Board Assembly

"License Key
Procedures".
Instrument Outer Case
Processor Assembly
removal
removal procedure.
removal procedure.
Power Supply Assembly
removal