Assembly Replacement Procedures
Front Frame Assembly
1. Refer to
2. Disconnect the cable A1W2 (25) from the LCD Inverter board (13) and
3. Disconnect the flex circuit A1W1 (24) from the front panel interface board
4. Remove the four screws (30) securing the display bracket (12) to the Front
Once the display assembly has been removed from the Front Frame Assembly, the
CAUTION
glass filter is no longer secured. DO NOT tip the assembly in such a manner that
would cause it to fall out of place, as this may cause injury and/or damage to the
glass.
5. Refer to
Figure 7-32
Inverter Board Removal
210
Figure
7-30. Ribbon cable W1 (26) must be separated from the
display by pulling up to separate the velcro that is used to adhere the cable to
the bracket.
unclip the cable from the cable clamps.
(9).
Frame Assembly. The display can now be removed from the Front Frame
Assembly.
Figure
7-32. Disconnect the backlight wires (2) from the inverter
board.
Chapter 7