Processor Assembly; Removal; Replacement - Agilent Technologies N9039A Service Manual

Rf preselector
Hide thumbs Also See for N9039A:
Table of Contents

Advertisement

Assembly Replacement Procedures

Processor Assembly

Processor Assembly

Removal

1. Refer to
2. The processor assembly can be removed from the chassis by pulling straight
Figure 7-12
Processor Assembly Removal

Replacement

1. Slide the processor assembly into the slot at the rear of the instrument and push
2. Refer to
178
Figure
7-12. Remove the six screws (1) attaching the processor
assembly to the chassis.
out the back. Use the two ejectors to pull the processor assembly out from the
chassis.
on the assembly to mate the connectors to the midplane assembly. Secure the
board with the ejectors.
Figure
7-12. Replace the six screws (1) that attach the processor
assembly to the chassis. Torque to 9 inch-pounds.