Performance Verification and Adjustment Software
Required Test Equipment
Table 10-1
Required Test Equipment for N9039A RF Preselector
Performance Tests
Instrument
3.5 mm (f) to Type-N (f)
3.5 mm (f) to 2.4 mm (f)
(2 required for 2.4 mm source)
BNC (m) to SMA(f)
Type-N (m) to 3.5 mm (f)
Type-N (f) to 2.4 mm (f)
Type-N (m) to 2.4 mm (f)
3.5 mm (f) to Type-N (m)
a. P = Performance Testing
b. Supported PSG models:
E8244A
E8254A
E8257C Option H31 or 540
E82578D
E8267D Option H31 or 540
E8267D
248
Critical Specifications
Frequency: DC to 18 GHz
VSWR: ≤ 1.08:1
Frequency: DC to 26.5 GHz
VSWR: ≤ 1.05:1
Frequency: DC to 1.3 GHz
VSWR: ≤ 1.13:1
Frequency: DC to 18 GHz
VSWR: ≤ 1.08:1
Frequency: DC to 18 GHz
VSWR: ≤ 1.08:1
Frequency: DC to 18 GHz
VSWR: ≤ 1.08:1
Shipped with the 8485 for adapting to the Power
Reference. Only to be used for power sensor cal.
Recommended
Alternative
Agilent Model
Agilent Model
Number
Number
1250-1745
11901B
1250-1700
1250-1744
11903B
11903D
08485-60005
a
Use
P
P
P
P
P
P
P
Chapter 10