Table 10-1
Required Test Equipment for N9039A RF Preselector
Performance Tests
Instrument
Cables
3.5 mm (m) to 3.5 mm (m)
(2 required)
Interconnect Cable
(for Type-N models)
Interconnect Cable
(for Option BAB models)
Interconnect Cable
(for 2.4 mm PSA)
Type-N
BNC
(4 required)
LAN
(3 required)
Filters
50 MHz Low Pass
300 MHz Low Pass
1.0 GHz Low Pass
Adapters
Type-N (m) to Type-N (m)
Type-N (m) to 3.5 mm (m)
(3 required)
Type-N (m) to BNC (f)
3.5 mm (f) to
3.5 mm (f)
(2 required for 3.5 mm source)
Chapter 10
Critical Specifications
Frequency: 9 kHz to 1.0 GHz
Length: ≤ 92 cm (36 in)
Insertion Loss: ~2 dB
VSWR: 9 kHz to 1.0 GHz: ≤ 1.25:1
Type-N (m) to Type-N (m)
3.5 mm (f) to 3.5 mm (f)
3.5 mm (f) to 2.4 mm (f)
Frequency: 9 kHz to 1.0 GHz
Precision Type-N (m), both ends
62 cm (24 in.)
VSWR: ≤ 18 GHz: 1.4:1
Insertion Loss: 1.5 dB
Frequency: 9 kHz to 30 MHz
50 Ω Coax BNC (m), both ends
120 cm (48 in.)
Shielded CAT5
Cutoff Frequency: 50 MHz
Rejection at 65 MHz: > 40 dB
Rejection at 75 MHz: > 60 dB
Insertion Loss: ~1 dB
VSWR: ≤ 1.5:1
BNC (m) to BNC (f)
Cutoff Frequency: 300 MHz
Rejection at > 435 MHz: > 45 dB
VSWR: ≤ 1.5:1
BNC (m, f)
Cutoff frequency: 1.0 GHz
Rejection at > 13 MHz: > 50 dB
Frequency: DC to 18 GHz
VSWR: ≤ 1.13:1
Frequency: DC to 18 GHz
VSWR: ≤ 1.08:1
Frequency: DC to 1.3 GHz
VSWR: ≤ 1.13:1
Frequency: DC to 26.5 GHz
VSWR: ≤ 1.05:1
Performance Verification and Adjustment Software
Required Test Equipment
Recommended
Alternative
Agilent Model
Agilent Model
Number
Number
8120-4921
11500E
N9039-21315
N9039-21316
N9039-21319
11500C
10503A
8121-1597
0955-0306
Telonic Berkeley
TLA 50-5AB2
0955-0455
Telonic Berkeley
TLP 300-4AB4
0955-0487
RLC Electronics
L-1621
1250-1475
1250-1743
1250-1476
83059B
1250-1749
a
Use
P
P
P
P
P
P
P
P
P
P
P
P
P
P