Agilent Technologies N9039A Service Manual page 245

Rf preselector
Hide thumbs Also See for N9039A:
Table of Contents

Advertisement

Table 10-1
Required Test Equipment for N9039A RF Preselector
Performance Tests
Instrument
Microwave Source 2
(only required for Third Order
Intermodulation)
RF Source 3
Function Generator 1
Network Analyzers
Network Analyzer 1
Network Analyzer 2
Reflection Test Set
Calibration Kit 1
Calibration Kit 2
Chapter 10
Critical Specifications
Frequency: 10 MHz to 26.5 GHz
Frequency Resolution: 1 Hz
Harmonic level: < −40 dBc
Amplitude range: −20dBm to +13
Amplitude resolution: 0.02
Harmonic level:
< 2 GHz, < -30 dBc
≥ 2 GHz & < 26.5 GHz, < -50 dBc
VSWR: < 20 GHz: 1.6:1
≤ 31 GHz: 1.8:1
Frequency: 100 kHz to 1.0 GHz
Amplitude = 0 dBm
Frequency: 9 kHz to 30 MHz
Amplitude Resolution: 0.1 mv
Harmonic Distortion: -35 dBc
Frequency: 30 MHz to 1.0 GHz
Directivity: 48 dB
Source Match: 40 dB
Reflection Tracking: ±0.003 dB
Frequency: 9 kHz to 30 MHz
Directivity: 40 dB
Source Match: 25 dB
Reflection Tracking: ±0.83 dB
Used with 4395A
50 ohm
Type-N (f)
Frequency: 9 kHz to 1.0 GHz
50 ohm
3.5 mm (m)
Frequency: 9 kHz to 1.0 GHz
Performance Verification and Adjustment Software
Required Test Equipment
Recommended
Alternative
Agilent Model
Agilent Model
Number
Number
b
83630A/B,
PSG
83640A/B,
83650A/B
(Option 001,008)
E8241A, E8244A,
E8251A, E8254A,
E8247C, E8257C
(Opt. H31 or 540,
1E1, 1EA)
E8247C, E8257C
(Opt. 1E1, 1EA,
520)
E8257D
(Options 1EA, 567)
E8267C
(Option 520)
E8267D
(Option 544 or 532
or 520)
b
PSG
33250A
N5230A
8753E
4395A
87512A
85032B
85032E.
85032F
85033E
a
Use
P
P
P
P
P
P
245