Power Supply assembly. This is used to frequency modulate the power supply
switching frequency for the purpose of lowering any power supply related
interference.
If for some reason the level of the power supply related interference is higher than
normal this circuitry, or the A6 Power Supply assembly, could be suspect.
Table 2-3
A7 Midplane Board LEDs and Test Points
LED
Description
+32V A
+32 Volt Analog
Supply
+15V A
+15 Volt Analog
Supply
+15V STBY
+15 Volt
Standby Supply
+12V D
+12 Volt Digital
Supply
+9V A
+9 Volt Analog
Supply
+5.1V A
+5.1 Volt
Analog Supply
+5.1V D
+5.1 Volt Digital
Supply
+5.1V STBY
+5.1 Volt
Standby Supply
+3.35V D
+3.35 Volt
Digital Supply
-7V A
-7 Volt Analog
Supply
-15V A
-15 Volt Analog
Supply
Fan P
Fan Positive
Voltage
+3.3V A
+3.3 Volt
Analog Supply
-5.2V A
-5.2 Volt Analog
Supply
Chapter 2
Reference
Regulated
Point
On
ACOM
A6
ACOM
A6
ACOM
A6
DCOM
A6
ACOM
A6
ACOM
A6
DCOM
A6
DCOM
A6
DCOM
A6
ACOM
A6
ACOM
A6
ACOM
A6
ACOM
A7
ACOM
A7
A7 Midplane Board Assembly
Expected Status
a
Power Off
LED
Test Point
(VDC)
Red
0
Off
Red
0
Off
+15 ± 1.0
Green
Green
Red
0
Off
Red
0
Off
Red
0
Off
Red
0
Off
+5.1 ± 0.75
Green
Green
Red
0
Off
Red
0
Off
Red
0
Off
Red
0
Off
Red
0
Off
Red
0
Off
Troubleshooting
Power On
LED
Test Point
(VDC)
+32 ± 2.0
+15 ± 1.0
+15 ± 1.0
+12 ± 1.0
+9 ± 1.0
+5.1 ± 0.75
+5.1 ± 0.75
+5.1 ± 0.75
+3.35 ± 0.75
-7 ± 1.0
-15 ± 1.0
9.2 - 14.5
+3.35 ± 0.75
-5.2 ± 0.75
65