Agilent Technologies N9039A Service Manual page 228

Rf preselector
Hide thumbs Also See for N9039A:
Table of Contents

Advertisement

Post-Repair Procedures
What You Will Find in This Chapter
7. When the Date and Time Settings screen come up as shown in
simply press the
have changed from what the end user has set it to be with an A5 Disk Drive
replacement.
Figure 8-11
Accept the Time Date Screen
8. After accepting the time and date screen the instrument will reboot a couple of
times. Just prior to the instrument rebooting for the last time a window will
appear on-screen as seen in
the instrument to shutdown and reboot.
Figure 8-12
Instrument Reboot Message
9. When the instrument has rebooted for the last time a window stating the fact
228
key to select the time that is displayed, as this will not
Enter
Figure 8-12
asking for you to wait for 2 minutes for
Figure 8-11
Chapter 8