Table 8-1
Post-Repair Testing Requirements (Continued)
Assembly
A21
Hardware Statistical Reset
Radiated Input
Overload Detector
Board
C-Band Bypass Abs Amp
R-Band Bypass Abs Amp
R-Band Step Gain Linearity
R-Band Nominal Gain
R-Band Abs Amp 0 and 10 dB
R-Band Abs Amp Versus Atten
A22
C-Band Bypass Abs Amp
Radiated Filter
R-Band Bypass Abs Amp
Board
R-Band Filter Tuning
R-Band Step Gain Linearity
R-Band Nominal Gain
A23
Hardware Statistical Reset
Conducted Input
Overload Detector
Board
C-Band Bypass Abs Amp
R-Band Bypass Abs Amp
C-Band Abs Amp 0 and 10 dB
R-Band Abs Amp Versus Atten
A24
C-Band Bypass Abs Amp
Conducted Filter
R-Band Bypass Abs Amp
Board
C-Band Abs Amp 0 and 10 dB
Sw1 or SW2
Hardware Statistical Reset
Switch Assemblies
C-Band Bypass Abs Amp
R-Band Bypass Abs Amp
C-Band Abs Amp 0 and 10 dB
R-Band Abs Amp 0 and 10 dB
a. Calibration Application Software is required to run the required performance tests and adjustments. For details go
to
http://www.agilent.com/find/calibrationsoftware.
b. The instrument application software must be updated to the latest version for certain repairs to keep the software
synchronized with the hardware. See
c. See
"BIOS Setup Changes"
d. Before replacing the A5 Disk Drive assembly see
e. See
"Programming Model and Serial Numbers"
f. See
"License Key Backup and Restoring"
g. Can be found on the Utilities menu of the Performance Verification and Adjustment Software.
Chapter 8
Adjustment to Perform
a
(in the order listed)
g
g
g
Chapter 9 , "Instrument
in this chapter.
"Calibration File Backup"
in this chapter.
in this chapter.
What You Will Find in This Chapter
Performance Verifications
to Perform After Adjustments
RF Preselector Alignment
Displayed Average Noise Level
Conducted Band Absolute Amplitude Accuracy
Radiated Band Absolute Amplitude Accuracy
Spurious Responses
Third Order Intercept
Conducted Band VSWR
Radiated Band VSWR
RF Preselector Alignment
Displayed Average Noise Level
Radiated Band Absolute Amplitude Accuracy
Spurious Responses
Third Order Intercept
Radiated Band VSWR
RF Preselector Alignment
Displayed Average Noise Level
Conducted Band Absolute Amplitude Accuracy
Spurious Responses
Third Order Intercept
Conducted Band VSWR
RF Preselector Alignment
Displayed Average Noise Level
Conducted Band Absolute Amplitude Accuracy
Spurious Responses
Third Order Intercept
Conducted Band VSWR
RF Preselector Alignment
Displayed Average Noise Level
Conducted Band Absolute Amplitude Accuracy
Radiated Band Absolute Amplitude Accuracy
Spurious Responses
Conducted Band VSWR
Radiated Band VSWR
Software".
in this chapter.
Post-Repair Procedures