Assembly Replacement Procedures
Conducted Filter Board Assembly
Replacement
1. Slide the Conducted Filter Board assembly (4) into the slot as shown in
Figure
7-21, using the extractors to secure the board to the connectors on the
motherboard assembly.
2. Replace W11. Torque to 10 inch-pounds.
3. Replace the top brace. Refer to the
4. Replace the instrument outer case. Refer to the
replacement procedure.
192
Top Brace
replacement procedure.
Instrument Outer Case
Chapter 7