Radiated Input Board Assembly

Removal

1. Remove the instrument outer case. Refer to the
procedure.
2. Remove the top brace. Refer to the
3. Refer to
Figure
7-21. Remove W7, W8, W9, W10, and W12.
4. The Radiated Input Board assembly (1) can be removed from the chassis by
pulling straight up using the extractors.

Replacement

1. Slide the Radiated Input Board assembly (1) into the slot as shown in
7-21, using the extractors to secure the board to the connectors on the

motherboard assembly.

2. Replace W7, W8, W9, W10, and W12. Torque to 10 inch-pounds.
3. Replace the top brace. Refer to the
4. Replace the instrument outer case. Refer to the
replacement procedure.
Chapter 7
Assembly Replacement Procedures
Radiated Input Board Assembly
Instrument Outer Case
Top Brace
removal procedure.
Top Brace
replacement procedure.
Instrument Outer Case
removal
Figure
195